[MRI of the knee joint: first results of a comparison of 0,2-T specialized system and 1,5-T high field strength magnet]

Rofo. 1995 May;162(5):390-5. doi: 10.1055/s-2007-1015905.
[Article in German]

Abstract

Purpose: Diagnostic accuracy and image quality of a specialised system for MR examinations of peripheral joints were evaluated.

Materials and methods: 20 patients with acute or chronic injuries of the knee were examined using a low-field MR system (0.2 T). For comparison, all patients were also studied with a 1.5 T high field strength magnet and all diagnoses were correlated with arthroscopic findings.

Results: We found compatible diagnostic accuracies (cruciate ligaments 90%, menisci 75-90%) and good image quality ratings for the low field system ("good" and "excellent" image quality in 83% of cases).

Conclusion: The low-field MR-system offers low-cost MR examinations of peripheral joints with good image quality and reliable diagnostic information.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Contusions / diagnosis
  • Evaluation Studies as Topic
  • Humans
  • Joint Dislocations / diagnosis
  • Knee Injuries / diagnosis*
  • Magnetic Resonance Imaging / methods*
  • Middle Aged
  • Sensitivity and Specificity