The effect of attenuation map, scatter energy window width, and volume of interest on the calibration factor calculation in quantitative 177Lu SPECT imaging: Simulation and phantom study

Phys Med. 2018 Dec:56:74-80. doi: 10.1016/j.ejmp.2018.11.009. Epub 2018 Nov 29.

Abstract

Purpose: The objective of this study was to evaluate the image degrading factors in quantitative 177Lu SPECT imaging when using both main gamma photopeak energies.

Methods: Phantom measurements with two different vials containing various calibrated activities in air or water were performed to derive a mean calibration factor (CF) for large and small volumes of interest (VOIs). In addition, Monte Carlo simulations were utilized to investigate the effect of scatter energy window width, scatter correction method, such as effective scatter source estimation (ESSE) and triple energy window (TEW), and attenuation map on the quantification of 177Lu.

Results: The measured mean CF using large and small VOIs in water was 4.50 ± 0.80 and 4.80 ± 0.72 cps MBq-1, respectively. Simulations showed a reference CF of 3.3 cps MBq-1 for the water-filled phantom considering all photons excluding scattered events. By using the attenuation map generated for 190 keV photons, the calculated CFs for 113 keV and 208 keV are 10% lower than by using the weighted mean energy of 175 keV for 177Lu. The calculated CF using the TEW correction was 17% higher than using the ESSE method for a water-filled phantom. However, our findings showed that an appropriate scatter window combination can reduce this difference between TEW and ESSE methods.

Conclusions: The present work implies that choosing a suitable width of scatter energy windows can reduce uncertainties in radioactivity quantification. It is suggested to generate the attenuation map at 113 keV and 208 keV, separately. Furthermore, using small VOIs is suggested in CF calculation.

Keywords: (177)Lu; Attenuation map; Calibration factor; SPECT/CT; Scatter fraction; Scatter window.

MeSH terms

  • Air
  • Calibration
  • Computer Simulation
  • Lutetium*
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radioisotopes*
  • Scattering, Radiation
  • Single Photon Emission Computed Tomography Computed Tomography / instrumentation
  • Single Photon Emission Computed Tomography Computed Tomography / methods*
  • Water

Substances

  • Radioisotopes
  • Water
  • Lutetium
  • Lutetium-177