Flow-mediated dilatation test using optoacoustic imaging: a proof-of-concept

Biomed Opt Express. 2017 Jun 23;8(7):3395-3403. doi: 10.1364/BOE.8.003395. eCollection 2017 Jul 1.

Abstract

Label-free multispectral optoacoustic tomography (MSOT) has recently shown superior performance in visualizing the morphology of human vasculature, especially of smaller vessels, compared to ultrasonography. Herein, we extend these observations towards MSOT interrogation of macrovascular endothelial function. We employed a real-time handheld MSOT scanner to assess flow-mediated dilatation (FMD), a technique used to characterize endothelial function. A data processing scheme was developed to quantify the dimensions and diameter changes of arteries in humans and determine wall distensibility parameters. By enabling high-resolution delineation of the blood-vessel wall in a cross-sectional fashion, the findings suggest MSOT as a capable alternative to ultrasonography for clinical FMD measurements.

Keywords: (000.1430) Biology and medicine; (110.5120) Photoacoustic imaging; (110.5125) Photoacoustics; (170.3880) Medical and biological imaging; (170.5380) Physiology.