Dual-source RF transmission in cardiac SSFP imaging at 3 T: systematic spatial evaluation of image quality improvement compared to conventional RF transmission

Clin Imaging. 2015 Mar-Apr;39(2):231-6. doi: 10.1016/j.clinimag.2014.11.007. Epub 2014 Nov 18.

Abstract

The purpose of this investigation was to systematically evaluate the spatial distribution of image quality improvement with dual-source radiofrequency (RF) transmission in cardiac steady-state free precession sequences at 3.0 T. Imaging with and without dual-source RF transmission was performed in 30 patients. Contrast-to-noise ratio for the left ventricular myocardium was significantly higher using dual-source RF transmission, but improvement was not uniformly distributed. The posterior myocardium showed significantly less contrast-to-noise ratio gain than all other cardiac regions. Signal-to-noise ratio increase was higher in the right than in the left ventricle. Subjective image quality was significantly enhanced by parallel RF transmission.