A matched-pair comparison of two different locking plates for valgus-producing medial open-wedge high tibial osteotomy: peek-carbon composite plate versus titanium plate

Knee Surg Sports Traumatol Arthrosc. 2015 Jul;23(7):2032-40. doi: 10.1007/s00167-014-2914-8. Epub 2014 Feb 22.

Abstract

Purpose: The first purpose of this study was to compare the clinical and radiographic outcome of two different locking plates used for valgus-producing medial open-wedge high tibial osteotomy (HTO). The second purpose was to histologically evaluate peek-carbon wear for biocompatibility.

Methods: Twenty-six consecutive patients undergoing open-wedge HTO using the first-generation PEEKPower HTO-Plate® (Group I) were matched with 26 patients after open-wedge HTO with the TomoFix™ plate (Group II). Clinical scores (visual analogue scale for pain, WOMAC, Lysholm score) were obtained preoperatively and at a minimum follow-up of 24 months postoperatively. Fixation stability was evaluated radiographically by comparing the medial proximal tibial angle (MPTA) and tibial slope 2 days after open-wedge HTO and after implant removal. Tissue samples of Group I were collected at the time of implant removal for histologic evaluation.

Results: Implant-related complications occurred in 15 % (n = 4) of Group I and 0 % of Group II. Out of them, 3 implant replacements were excluded from statistical analyses. After a final median follow-up of 25 months (range 24-31), the clinical scores in both groups showed significant improvements compared to preoperatively (visual analogue scale, WOMAC, Lysholm score; p < 0.001), without significant group differences (visual analogue scale, n.s.; WOMAC, n.s.; Lysholm score, n.s.). No significant differences between baseline and follow-up measurements for MPTA and tibial slope were observed within each group (MPTA: Gr. I, n.s.; Gr. II, n.s.; tibial slope: Gr. I, n.s.; Gr. II, n.s.) or between the two groups (MPTA, n.s.; tibial slope, n.s.). In histologic samples, CF PEEK abrasion did not induce inflammation or tissue necrosis.

Conclusion: The first-generation PEEKPower HTO-Plate® provided a higher rate of implant-related complications compared to the TomoFix™ plate at a minimum follow-up of 24 months after valgus-producing open-wedge HTO. Therefore, it is not recommended to use the first-generation PEEKPower HTO-Plate® in the clinical practice.

Level of evidence: III.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Benzophenones
  • Bone Plates* / adverse effects
  • Carbon
  • Device Removal
  • Equipment Design
  • Female
  • Humans
  • Ketones
  • Male
  • Matched-Pair Analysis
  • Middle Aged
  • Osteotomy / instrumentation*
  • Polyethylene Glycols
  • Polymers
  • Postoperative Complications
  • Radiography
  • Tibia / diagnostic imaging
  • Tibia / surgery*
  • Titanium

Substances

  • Benzophenones
  • Ketones
  • Polymers
  • polyetheretherketone
  • Polyethylene Glycols
  • Carbon
  • Titanium