Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose

Phys Med Biol. 1990 Sep;35(9):1211-9. doi: 10.1088/0031-9155/35/9/002.

Abstract

The IPSM report on the commissioning and routine testing of mammographic x-ray systems recommends that breast dose be specified as the mean dose to the glandular tissues within the breast and gives the size and compositions of a standard breast phantom for the comparison of doses. The dose to this standard breast phantom can be determined by measuring the incident air kerma to a Perspex phantom and applying appropriate multiplicative conversion factors. These conversion factors have been evaluated by Monte Carlo calculations for a wide range of mammographic x-ray spectra. Some factors are provided for a range of breast thicknesses to supplement existing tabulations. Results are also given for equivalent thicknesses of Perspex and breast tissue.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Female
  • Humans
  • Mammography / standards*
  • Models, Structural
  • Monte Carlo Method
  • Radiation Dosage*