MR visualization of the inner ear structures: comparison of 1.5 Tesla and 3 Tesla images

Rofo. 2004 Jan;176(1):17-20. doi: 10.1055/s-2004-814672.

Abstract

Aim of the study: To compare high resolution MRI examinations of inner ear structures at 1.5 T and at 3 T.

Method: Temporal bones were measured bilaterally in 3 healthy volunteers in a 1.5 T and in a 3 T MR-scanner using the respective one channel head coil (quadrature detection) of the manufacturer. The same steady-state gradient echo sequence (3D-CISS) was employed at a voxel size of 0.4 x 0.4 x 0.4 mm(3). The signal-to-noise ratio (SNR) was determined quantitatively.

Results: An SNR of 8 could be achieved for the measurements at 3 T in 7:37 min. The SNR at 3 T was, on average, a factor of 1.34 higher than that at 1.5 T despite the fact that the excitation angle had to be drastically reduced (alpha = 42 degrees instead of alpha = 70 degrees at 1.5 T) due to the limit of the specific absorption rate (SAR).

Discussion: The MR representation of the inner ear is clearly improved at 3 T. To obtain the same SNR at 1.5 T approximately the double measuring time would be required, connected with reduced patient comfort and an increased risk for a displacement of the head during the high resolution measurement.